• DocumentCode
    808633
  • Title

    Properties of Y2O2S:Tb X-Ray Intensifying Screens

  • Author

    Alves, R.V. ; Buchanan, R.A.

  • Author_Institution
    Lockheed Palo Alto Research Laboratory Palo Alto, California 94304
  • Volume
    20
  • Issue
    1
  • fYear
    1973
  • Firstpage
    415
  • Lastpage
    419
  • Abstract
    Experimental radiographic screens of Y2O2S:Tb, when using ordinary (blue-sensitive) medical x-ray film, are shown to have a speed three to four times that of commercial CaWO4, and about twice that of Gd2O2S:Tb in the medical x-ray energy range of 30 to 100 kVp. The resolutions in each case were comparable. Calculations of the relative speed of these screens are made from the x-ray attenuation coefficients, the cathodoluminescent efficiencies, and the spectral response of the film. These calculations agree quite well with the observed values. Resolution data are presented in terms of the square wave response function for experimental and commercial radiographic screens. Results of exposures made with an experimental x-ray film sensitized for the green spectral region with Y2O2S:Tb and Gd2O2S:Tb radiographic intensifying screens are presented.
  • Keywords
    Atomic measurements; Attenuation; Detectors; Electromagnetic wave absorption; Energy resolution; Filters; Laboratories; Phosphors; Radiography; Wideband;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1973.4326938
  • Filename
    4326938