• DocumentCode
    808778
  • Title

    Collimated Beam Scanning of Pure Germanium Detectors

  • Author

    Baertsch, R.D.

  • Author_Institution
    General Electric Corporate Research and Development Schenectady, New York 12301
  • Volume
    20
  • Issue
    1
  • fYear
    1973
  • Firstpage
    488
  • Lastpage
    493
  • Abstract
    The design of collimated sources of Am241, Co57, and Fe55 suitable for scanning germanium detectors is given. Am241 scans taken through the p contact on a p-type detector are given for detectors with and without radial resistivity gradients. The data obtained from these scans are used to calculate the radial dependence of depletion layer width and an "effective" carrier concentration. The results obtained indicate that, with the surface treatment used, the surface is accumulated on p-type detectors. At voltages below full depletion, an extra "peak" associated with slow pulses originating in the undepleted region is described. The number of these slow pulses depends on the type of back contact that is used.
  • Keywords
    Apertures; Collimators; Conductivity; Etching; Face detection; Germanium; Impurities; Radiation detectors; Research and development; Surface treatment;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1973.4326952
  • Filename
    4326952