• DocumentCode
    809180
  • Title

    Uncertainty specification for data acquisition (DAQ) devices

  • Author

    Braudaway, David W.

  • Author_Institution
    Sandia Nat. Labs., Albuquerque, NM, USA
  • Volume
    55
  • Issue
    1
  • fYear
    2006
  • Firstpage
    74
  • Lastpage
    78
  • Abstract
    Specification of uncertainty has historically been done by a variety of methods with differences in results, especially for instruments and standards that achieve small uncertainty values. A relatively simple but practical approach is used in newly published standard IEC 62008 1st Edition, "Performance characteristics and calibration methods for digital data acquisition systems and relevant software." The standard currently covers only relevant manufacturer\´s specifications for direct current (dc) data acquisition (DAQ) devices. These are plug-in cards without the associated computer for readout and control or power supply. Rather than complete instruments, the cards require an associated computer for readout and control. The history of uncertainty is reviewed; some popular current techniques for specifying uncertainty are compared. The components of uncertainty chosen for DAQ plug-in cards are reviewed, and their effects are discussed.
  • Keywords
    IEC standards; calibration; data acquisition; measurement standards; measurement uncertainty; calibration methods; digital data acquisition systems; direct current data acquisition devices; standard IEC 62008; uncertainty specification; Calibration; Data acquisition; IEC standards; Instruments; Manufacturing; Software performance; Software standards; Software systems; Standards publication; Uncertainty; Calibration; coverage factor; expanded uncertainty DAQ; statistics; uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2005.862014
  • Filename