    A method for the calibration of harmonic analysers using signals containing fluctuating harmonics in support of IEC61000-3-2

  • Author

    Wright, P.S.

  • Author_Institution
    Nat. Phys. Lab., Teddington, UK
  • Volume
    152
  • Issue
    3
  • fYear
    2005
  • fDate
    5/6/2005 12:00:00 AM
  • Firstpage
    103
  • Lastpage
    109
  • Abstract
    The motivation for the calibration of harmonic analysers under fluctuating harmonic conditions is considered. An overview of the prescribed design of IEC61000-4-7 harmonic analysers is given. A class of signal is presented that can be used to calibrate analysers at the ´maximum´ and ´average´ limits as defined in the international standard. The inherent variation in results identified as a result distribution due to variable phase (RDVP) is considered together with the consequences of the RDVP for the use and calibration of these harmonic analysers. A method to fit analyser results to the RDVP, based on multiple phase shifted harmonic modulators is presented, such that analyser performance can be assessed. Sample results are given, showing significant errors on commercial devices, despite their apparent correct performance under steady-state conditions.
  • Keywords
    IEC standards; calibration; harmonic analysis; spectral analysers; IEC61000-3-2; RDVP; calibration; fluctuating harmonics; harmonic analysers; international standard; multiple phase shifted harmonic modulators; result distribution due to variable phase;
  • fLanguage
    English
  • Journal_Title
    Science, Measurement and Technology, IEE Proceedings -
  • Publisher
    iet
  • ISSN
    1350-2344
  • Type

    jour

  • DOI
    10.1049/ip-smt:20049028
  • Filename
    1432558