Title :
Recent Advances in Gas Scintillation Proportional Counters
Author :
Palmer, H.E.
Author_Institution :
Battelle Pacific Northwest Laboratory Richland, Washington 99352
Volume :
22
Issue :
1
fYear :
1975
Firstpage :
100
Lastpage :
103
Abstract :
Various geometrical configurations for gas scintillation proportional counters have been investigated in order to determine which is best for use in a large volume, high efficiency counter for measuring low energy gamma and x-rays. A xenon filled counter having a rod anode inside a cylindrical cathode appears to provide the best configuration for providing a uniform field and the best resolution over the total volume of the counter. The details of construction and operating characteristics of various shaped counters are described.
Keywords :
Anodes; Cathodes; Energy resolution; Laboratories; Photomultipliers; Radiation detectors; Wire; X-ray detection; X-ray detectors; Xenon;