• DocumentCode
    817606
  • Title

    The influence of nonlinear fields on miniature hydrophone calibration using the planar scanning technique

  • Author

    Corbett, Scott S., III

  • Author_Institution
    Gen. Elec. Co., Rancho Cordova, CA, USA
  • Volume
    35
  • Issue
    2
  • fYear
    1988
  • fDate
    3/1/1988 12:00:00 AM
  • Firstpage
    162
  • Lastpage
    167
  • Abstract
    Miniature polyvinylidene fluoride (PVDF) hydrophones used in determining the power and intensity output of ultrasonic fields, including those radiated from diagnostic ultrasound equipment, were calibrated under a variety of field conditions using the planar scanning technique. A diagnostic B-scan piston-type transducer was used as a source, and the output intensity (spatial-peak pulse-average, or SPPA) was varied from 2 to 30 W/cm/sup 2/ while the total power was kept constant. The higher-intensity waveforms were significantly nonlinear in the focal region of the source. When the lateral beam profiles of the source (as measured by the hydrophone to be calibrated) were determined by positive-peak-detecting the ultrasonic pulse, the calibrated pressure sensitivity of the hydrophone systematically decreased as the field became progressively more nonlinear. When the beam profiles were measured using the pulse intensity integral, no systematic dependence of the calibration on field linearity was noted. These results imply that measured values of power and intensity of ultrasound diagnostic equipment may be dependent on the methodology utilized to map the lateral beam profiles of the transducer being measured, and the extent of nonlinear effects in the field under characterization.<>
  • Keywords
    calibration; hydrophones; B-scan piston-type transducer; PVDF; diagnostic ultrasound; intensity; lateral beam profiles; miniature hydrophone calibration; nonlinear fields; planar scanning; power; pulse intensity integral; ultrasonic fields; Acoustic beams; Calibration; Force measurement; Power measurement; Pressure measurement; Pulse measurements; Sonar equipment; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.4166
  • Filename
    4166