• DocumentCode
    820221
  • Title

    Investigation of coil geometry for transcutaneous energy transmission for artificial heart

  • Author

    Matsuki, H. ; Shiiki, M. ; Murakami, K. ; Yamamoto, T.

  • Author_Institution
    Dept. of Electr. Eng., Tohoku Univ., Sendai, Japan
  • Volume
    28
  • Issue
    5
  • fYear
    1992
  • fDate
    9/1/1992 12:00:00 AM
  • Firstpage
    2406
  • Lastpage
    2408
  • Abstract
    Transcutaneous energy transmission for a totally implantable artificial heart system is considered, and noncontact electric energy transmission by means of spiral coils with amorphous magnetic fiber is reported. Coil design for higher transmitting efficiency and a smaller temperature rise of the coil has been investigated. An experiment in vitro showed that temperature rise will be reduced within 3°C
  • Keywords
    artificial organs; coils; amorphous magnetic fiber; artificial heart; coil geometry; noncontact electric energy transmission; spiral coils; transcutaneous energy transmission; Artificial heart; Coils; Equations; Equivalent circuits; Geometry; Impedance; Iron; Resonance; Skin; Temperature;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.179507
  • Filename