Title :
A Microprocessor-Assisted Calibration for a Remote Working Level Monitor
Author :
McDowell, William P. ; Keefe, Donald J. ; Groer, Peter G. ; Witek, Richard T.
Author_Institution :
Argonne National Laboratory, Argonne, Illinois 60439
Abstract :
A method is described for calibrating a Remote Working Level Monitor, an instrument which measures Working Level and Rn-daughter concentrations in the atmosphere. The method makes use of a microprocessor to calculate beta efficiencies for RaB and RaC from the counts accumulated in the RaA, Ra(B + C) and RaC´ channels of the instrument. Both the alpha spectroscopic and total-alpha methods are used to determine the Rn-daughter concentrations. These methods require the processor to solve systems of linear equations with several unknowns. No assumptions about Rn-daughter equilibrium are made.
Keywords :
Atmosphere; Atmospheric measurements; Calibration; Detectors; Equations; Instruments; Laboratories; Microprocessors; Remote monitoring; Time measurement;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1977.4328756