DocumentCode
830377
Title
Long Radiation Detector System for Beam Loss Monitoring
Author
Balsamo, J. ; Fewell, N.M. ; Klein, J.D. ; Witkover, R.L.
Author_Institution
Brookhaven National Laboratory Upton, New York 11973
Volume
24
Issue
3
fYear
1977
fDate
6/1/1977 12:00:00 AM
Firstpage
1807
Lastpage
1809
Abstract
This paper describes the Long Radiation Monitor (LRM) system installed at the 200 MeV linac at Brookhaven National Laboratory. This system allows observation of both the spatial and temporal character of the losses in the linac and its transport lines. An array of large diameter gas filled coaxial cables are used as extended ion chambers to detect the losses. The output signals are available as a histogram, video waveforms, and numerical data via the computer. A fast beam interrupt is also provided. The detector characteristics and details of the processing electronics are presented. Results of studies of longitudinal, steering and focusing losses are described.
Keywords
Coaxial cables; Electrons; Histograms; Isotopes; Laboratories; Linear particle accelerator; Protons; Pulse measurements; Radiation detectors; Radiation monitoring;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1977.4329094
Filename
4329094
Link To Document