• DocumentCode
    831385
  • Title

    NIST-NPL interlaboratory pulse measurement comparison

  • Author

    Paulter, N.G. ; Smith, Andrew J A ; Larson, D.R. ; Souders, T.M. ; Roddie, Alan G.

  • Author_Institution
    Electr.ity Div., Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • Volume
    52
  • Issue
    6
  • fYear
    2003
  • Firstpage
    1825
  • Lastpage
    1833
  • Abstract
    A comparison of the pulse parameter values obtained from the pulse measurement services of the National Institute of Standards and Technology, USA, and the National Physical Laboratory, U.K., was performed. The comparison was based on the pulse parameters of amplitude, transition duration, overshoot, and undershoot (preshoot). The parameter comparison was applied to raw (measured) waveforms, corrected waveforms (if applicable), and reconstructed waveforms. The results of the comparison show that the pulse parameter values for both national laboratories are within published uncertainties.
  • Keywords
    calibration; measurement uncertainty; pulse generators; pulse measurement; signal sampling; transient response; amplitude; corrected waveforms; equivalent-time sampling oscilloscopes; high-speed pulse generators; impulse responses; interlaboratory pulse measurement comparison; measurement uncertainties; national laboratories; overshoot; preshoot; pulse measurement services; pulse parameter values; raw waveforms; reconstructed waveforms; transition duration; undershoot; waveform processing; Attenuation measurement; Computer errors; Error correction; Gain measurement; Laboratories; NIST; Performance evaluation; Pulse generation; Pulse measurements; Time measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2003.820456
  • Filename
    1246556