• DocumentCode
    833426
  • Title

    A High Rate Amplifier-Digitizer System for Liquid Argon Calorimeters

  • Author

    Droege, T.F. ; Lobkowicz, F. ; Fukushima, Y.

  • Author_Institution
    Fermi National Accelerator Laboratory P. O. Box 500, Batavia Ill.
  • Volume
    25
  • Issue
    1
  • fYear
    1978
  • Firstpage
    687
  • Lastpage
    691
  • Abstract
    A low cost charge amplifier for a liquid argon photon detector and a new method for pulse height analysis are described. This scheme is suitable for high energy photon detection with high counting rate. Samples of preamplifier output are taken just before and just after the arrival of the charge from the detector. The difference of these samples provides a stable pedestal and rejects low frequency noise. Short two-pulse resolving time (¿200ns) is achieved.
  • Keywords
    Argon; Capacitance; Detectors; Energy resolution; Low-frequency noise; Low-noise amplifiers; Pulse amplifiers; Pulse measurements; Pulse transformers; Strips;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1978.4329393
  • Filename
    4329393