DocumentCode
833426
Title
A High Rate Amplifier-Digitizer System for Liquid Argon Calorimeters
Author
Droege, T.F. ; Lobkowicz, F. ; Fukushima, Y.
Author_Institution
Fermi National Accelerator Laboratory P. O. Box 500, Batavia Ill.
Volume
25
Issue
1
fYear
1978
Firstpage
687
Lastpage
691
Abstract
A low cost charge amplifier for a liquid argon photon detector and a new method for pulse height analysis are described. This scheme is suitable for high energy photon detection with high counting rate. Samples of preamplifier output are taken just before and just after the arrival of the charge from the detector. The difference of these samples provides a stable pedestal and rejects low frequency noise. Short two-pulse resolving time (¿200ns) is achieved.
Keywords
Argon; Capacitance; Detectors; Energy resolution; Low-frequency noise; Low-noise amplifiers; Pulse amplifiers; Pulse measurements; Pulse transformers; Strips;