• DocumentCode
    836348
  • Title

    Microcomputer Based Dual Energy Photon Absorptiometric Bone Mineral Analyzer (VCH)

  • Author

    Vogel, John M. ; Cline, James W. ; Harrison, James F. ; Ulloa, G.A. ; McDonald, R.J.

  • Author_Institution
    Dominican Santa Cruz Hospital
  • Volume
    26
  • Issue
    1
  • fYear
    1979
  • Firstpage
    576
  • Lastpage
    582
  • Abstract
    Measurement of mineral content in trabecular bone with reproducible precision can have significant effect on patient management with osteopenic conditions. Commercial instruments measure single cross sections of the upper extremity which, unless uniform bone sections are measured, produce large errors. Forearm bones, that are primarily cortical and therefore less reflective of rapid bone turnover, are usually measured. Utilizing experience gained with rectilinear scanning for large area mineral estimation on Apollo and Skylab astronauts, we designed hardware and software that is compact, can use single or dual photon sources in a rectilinear fashion with precise repositioning, scan in a vertical or horizontal plane, perform on-line data storage and reduction, and provide continuous visual evaluation of the data profile.
  • Keywords
    Cancellous bone; Content management; Extraterrestrial measurements; Extremities; Hardware; Instruments; Microcomputers; Minerals; Software design; Software performance;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1979.4329693
  • Filename
    4329693