DocumentCode
836709
Title
Gas proportional scintillation counters for the μp-Lamb shift experiment
Author
Veloso, J.F.C.A. ; Lopes, J.A.M. ; Conde, C.A.N. ; Fernandes, L.M.P. ; Freitas, E.D.C. ; Huot, O. ; Knowles, P. ; Kottmann, F. ; Mulhauser, F. ; Dos Santos, J.M.F. ; Taqqu, D.
Author_Institution
Dept. de Fisica, Coimbra Univ., Portugal
Volume
49
Issue
3
fYear
2002
fDate
6/1/2002 12:00:00 AM
Firstpage
899
Lastpage
906
Abstract
Large-area 1.9-keV X-ray detectors operating in magnetic fields up to 5 T are required for the μp-Lamb shift experiment. Xenon gas proportional scintillation counters provide high detection efficiency together with good energy and timing resolutions. Three prototypes with alternative vacuum ultraviolet photosensors of the xenon scintillation light are explored and discussed: a CsI-coated microstrip plate either integrated within the xenon envelope or in a separate chamber in a P-10 atmosphere and an avalanche photodiode integrated within the xenon envelope.
Keywords
X-ray detection; gas scintillation detectors; proportional counters; xenon; CsI-coated microstrip plate; Xe; Xe gas proportional scintillation counters; avalanche photodiode; magnetic field; muon-p-Lamb shift experiment; vacuum ultraviolet photosensors; Atmosphere; Energy resolution; Magnetic fields; Microstrip; Prototypes; Scintillation counters; Solid scintillation detectors; Timing; X-ray detectors; Xenon;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2002.1039586