• DocumentCode
    836769
  • Title

    Comparison of neutron scintillation detectors with a 3He proportional counter for the Spallation Neutron Source (SNS)

  • Author

    Engels, R. ; Kemmerling, G. ; Rongen, H. ; Schelten, J. ; Cooper, R.

  • Author_Institution
    Zentrallabor fur Elektron., Forschungszentrum Julich GmbH, Germany
  • Volume
    49
  • Issue
    3
  • fYear
    2002
  • fDate
    6/1/2002 12:00:00 AM
  • Firstpage
    923
  • Lastpage
    925
  • Abstract
    The performance of scintillation detectors with the advanced scintillator material 6Li 158Gd (11BO3)3 was compared with conventional 3He proportional counters at the IPNS spallation source at the Argonne National Laboratory, Argonne, IL. The complex neutron and gamma flux was measured with both detector types simultaneously in time-of-flight mode. Initial experiments demonstrated that both detectors have a sensitivity of about 60% for thermal neutrons. In addition, both detectors exhibited the same low gamma sensitivity.
  • Keywords
    gamma-ray detection; helium-3 counters; neutron detection; neutron sources; solid scintillation detectors; 3He proportional counter; He; LiGd(BO3)3; SNS; Spallation Neutron Source; gamma flux; neutron flux; neutron scintillation detectors; Gamma ray detection; Gamma ray detectors; Helium; Laboratories; Neutrons; Radiation detectors; Scattering; Scintillation counters; Sea measurements; Solid scintillation detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2002.1039590
  • Filename
    1039590