• DocumentCode
    838040
  • Title

    The Use of Compact Cyclotrons for Producing Fast Neutrons for Therapy in a Rotatable Isocentric Gantry

  • Author

    Tom, James L.

  • Author_Institution
    The Cyclotron Corporation Berkeley, California 94710
  • Volume
    26
  • Issue
    2
  • fYear
    1979
  • fDate
    4/1/1979 12:00:00 AM
  • Firstpage
    2294
  • Lastpage
    2298
  • Abstract
    This paper discusses the design and performance of a fast neutron therapy system which transports a cyclotron beam onto a beryllium target mounted in a rotatable isocentric gantry structure. Fast neutrons produced by either the proton or deuteron on beryllium reaction are directed towards the isocenter with appropriate shielding, collimation and dosimetry to control and monitor the radiation delivered onto a defined patient area. The beam optics and shielding considerations of such a radiotherapy apparatus are discussed. This paper also presents the dosimetric properties of neutron beams from different proton energies on beryllium and lithium targets.
  • Keywords
    Cyclotrons; Dosimetry; Medical treatment; Neutrons; Optical beams; Optical collimators; Particle beams; Patient monitoring; Protons; Radiation monitoring;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1979.4329861
  • Filename