DocumentCode
838040
Title
The Use of Compact Cyclotrons for Producing Fast Neutrons for Therapy in a Rotatable Isocentric Gantry
Author
Tom, James L.
Author_Institution
The Cyclotron Corporation Berkeley, California 94710
Volume
26
Issue
2
fYear
1979
fDate
4/1/1979 12:00:00 AM
Firstpage
2294
Lastpage
2298
Abstract
This paper discusses the design and performance of a fast neutron therapy system which transports a cyclotron beam onto a beryllium target mounted in a rotatable isocentric gantry structure. Fast neutrons produced by either the proton or deuteron on beryllium reaction are directed towards the isocenter with appropriate shielding, collimation and dosimetry to control and monitor the radiation delivered onto a defined patient area. The beam optics and shielding considerations of such a radiotherapy apparatus are discussed. This paper also presents the dosimetric properties of neutron beams from different proton energies on beryllium and lithium targets.
Keywords
Cyclotrons; Dosimetry; Medical treatment; Neutrons; Optical beams; Optical collimators; Particle beams; Patient monitoring; Protons; Radiation monitoring;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1979.4329861
Filename
4329861
Link To Document