Title
Radiation response characteristics of amorphous silicon arrays for megavoltage radiotherapy imaging
Author
Antonuk, L.E. ; Yorkston, J. ; Huang, W. ; Boudry, J. ; Morton, E.J. ; Longo, M.J. ; Street, R.A.
Author_Institution
Dept. of Radiat. Oncology, Michigan Univ. Med. Center, Ann Arbor, MI, USA
Volume
39
Issue
4
fYear
1992
fDate
8/1/1992 12:00:00 AM
Firstpage
1069
Lastpage
1073
Abstract
Signal measurements for a variety of pixelated a-Si:H imaging arrays irradiated by megavoltage beams are reported. Data are presented for 6-, 10-, and 15-MV beam energies for screens containing gadolinium-oxysulfide and calcium tungstate. These results are quantitative radiation signal sizes reported for such arrays. The initialization times of these arrays are reported. The performance of arrays of various sizes and pitches with respect to these measured values is compared. These results are part of an ongoing effort to develop flat-panel, real-time gamma and X-ray imagers
Keywords
biomedical equipment; hydrogen; radiation therapy; semiconductor counters; silicon; 6 to 15 MV; X-ray imagers; amorphous Si:H arrays; initialization times; megavoltage beams; megavoltage radiotherapy imaging; real time gamma imaging; Amorphous silicon; Biomedical imaging; Calcium; Collaboration; Image quality; Medical diagnostic imaging; Optical imaging; Pixel; Readout electronics; X-ray imaging;