• DocumentCode
    841930
  • Title

    Ultrasound holographic B-scan imaging

  • Author

    Ylitalo, Juha ; Alasaarela, Esko ; Koivukangas, John

  • Author_Institution
    Dept. of Electr. Eng., Oulu Univ., Finland
  • Volume
    36
  • Issue
    3
  • fYear
    1989
  • fDate
    5/1/1989 12:00:00 AM
  • Firstpage
    376
  • Lastpage
    383
  • Abstract
    An ultrasound holographic B-scan (UHB) imaging apparatus comprising a minicomputer system, a data acquisition unit, and a special 64-element UHB transducer has been developed. Simulation studies and real experiments with a tissue-equivalent phantom show that lateral and longitudinal resolution (-6 dB) of about 1 mm was achieved in the entire image. Furthermore, results from clinical evaluation, including diagnostic and neurosurgical imaging, suggest that the UHB imaging method is operational and has some special advantages in patient diagnosis. Theoretically, the addition of phase information to the ultrasound images can result in enhanced tissue characterization, which is extremely important in tumor diagnosis and treatment.<>
  • Keywords
    acoustic holography; acoustic imaging; acoustic signal processing; biomedical ultrasonics; patient diagnosis; 64-element UHB transducer; acoustic imaging; biomedical ultrasound; data acquisition unit; holographic B-scan imaging; minicomputer system; phase information; resolution; tissue characterization; tissue-equivalent phantom; treatment; tumor diagnosis; ultrasound; Data acquisition; Holography; Image resolution; Imaging phantoms; Medical diagnosis; Microcomputers; Neoplasms; Neurosurgery; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.19178
  • Filename
    19178