• DocumentCode
    845384
  • Title

    Accuracy of measurements of the quantized Hall resistivity by a direct current comparator type potentiometer: calibration using a Josephson potentiometer

  • Author

    Inagaki, Katsuya ; Sakamoto, Yasuhiko ; Endo, Tadashi

  • Author_Institution
    Electrotech. Lab., Ibaraki, Japan
  • Volume
    38
  • Issue
    2
  • fYear
    1989
  • fDate
    4/1/1989 12:00:00 AM
  • Firstpage
    276
  • Lastpage
    278
  • Abstract
    The accuracy of measurements of the quantized Hall resistivity (QHR) by a direct-current-comparator (DCC)-type potentiometer is estimated based on a linearity calibration of the potentiometer using a Josephson potentiometer. The power coefficient contribution to the nonlinearity of the DCC potentiometer was found to be 0.15±0.02 p.p.m./(100 mV)2 at 18.5±0.5°C in 1985 and 0.21±0.03 p.p.m./(100 mV)2 at 20.5±0.5°C in 1988. The possibility of accurate measurements of the ratio QHR/R STD with uncertainties less than 0.05 p.p.m. by the DCC potentiometer is discussed
  • Keywords
    calibration; constants; electric resistance measurement; potentiometers; quantum Hall effect; Josephson potentiometer; calibration; constants; direct current comparator; nonlinearity; power coefficient; quantized Hall resistivity; Calibration; Conductivity; Current measurement; Electrical resistance measurement; Frequency; Josephson junctions; Linearity; Measurement standards; Microwave antenna arrays; Potentiometers;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.192287
  • Filename
    192287