DocumentCode
84716
Title
Robot-Guided Open-Loop Insertion of Skew-Line Needle Arrangements for High Dose Rate Brachytherapy
Author
Garg, Adesh ; Siauw, Timmy ; Berenson, Dmitry ; Cunha, J. Adam M. ; I-Chow Hsu ; Pouliot, Jean ; Stoianovici, D. ; Goldberg, K.
Author_Institution
Dept. of Ind. Eng. & Oper. Res., Univ. of California, Berkeley, Berkeley, CA, USA
Volume
10
Issue
4
fYear
2013
fDate
Oct. 2013
Firstpage
948
Lastpage
956
Abstract
We present a study in human-centered automation that has potential to reduce patient side effects from high dose rate brachytherapy (HDR-BT). To efficiently deliver radiation to the prostate while minimizing trauma to sensitive structures such as the penile bulb, we modified the Acubot-RND 7-axis robot to guide insertion of diamond-tip needles into desired skew-line geometric arrangements. We extend and integrate two algorithms: Needle Planning with Integer Programming (NPIP) and Inverse Planning with Integer Programming (IPIP) to compute skew-line needle and dose plans. We performed three physical experiments with anatomically correct phantom models to study performance: two with the robot and one control experiment with an expert human physician (coauthor Hsu) without the robot. All were able to achieve needle arrangements that meet the RTOG-0321 clinical dose objectives with zero trauma to the penile bulb. We analyze systematic and random errors in needle placement; total RMS error for the robot system operating without feedback ranged from 2.6 to 4.3 mm, which is comparable to the RMS error of 2.7 mm obtained in an earlier study for PPI-BT treatment using a robot with 3D ultrasound feedback.
Keywords
brachytherapy; integer programming; medical robotics; needles; Acubot-RND 7-axis robot; RMS error; RTOG-0321 clinical dose objectives; high dose rate brachytherapy; inverse planning integer programming algorithm; needle arrangements; needle planning integer programming algorithm; prostate; robot-guided open loop insertion; skew line needle arrangements; Biomedical applications of radiation; Brachytherapy; Medical robotics; Needles; Prostate cancer; Surgery; Brachytherapy; health care; needle insertion; prostate cancer; radiation; robot; robot assisted surgery; steerable needles;
fLanguage
English
Journal_Title
Automation Science and Engineering, IEEE Transactions on
Publisher
ieee
ISSN
1545-5955
Type
jour
DOI
10.1109/TASE.2013.2276940
Filename
6579776
Link To Document