  • Title

    Techniques for X-Ray Fluorescence Tomography

  • Author

    Patton, J.A. ; Price, R.R. ; Pickens, D.R. ; Brill, A.B.

  • Author_Institution
    Vanderbilt Medical Center, Nashville, Tennessee
  • Volume
    27
  • Issue
    1
  • fYear
    1980
  • Firstpage
    421
  • Lastpage
    424
  • Abstract
    New techniques are being explored to improve the sensitivity, resolution and diagnostic information content of currently available x-ray fluorescence techniques for imaging the stable iodine distribution within the thyroid gland. A new source and detector collimation system has been fabricated for a 9-element array of high purity germanium detectors to provide simultaneous longitudinal section images of uptake (Tc-99m pertechnetate) and iodine content. An added feature of this new system is the capability of performing stationary counts to plot vertical profiles of iodine content through any portion of the gland. This new system should be very useful for studies designed to differentiate between malignant and benign thyroid nodules on the basis of iodine content.
  • Keywords
    Collimators; Detectors; Fluorescence; Glands; High-resolution imaging; Image resolution; Optical imaging; Sensor arrays; Tomography; X-ray imaging;