DocumentCode
848541
Title
Techniques for X-Ray Fluorescence Tomography
Author
Patton, J.A. ; Price, R.R. ; Pickens, D.R. ; Brill, A.B.
Author_Institution
Vanderbilt Medical Center, Nashville, Tennessee
Volume
27
Issue
1
fYear
1980
Firstpage
421
Lastpage
424
Abstract
New techniques are being explored to improve the sensitivity, resolution and diagnostic information content of currently available x-ray fluorescence techniques for imaging the stable iodine distribution within the thyroid gland. A new source and detector collimation system has been fabricated for a 9-element array of high purity germanium detectors to provide simultaneous longitudinal section images of uptake (Tc-99m pertechnetate) and iodine content. An added feature of this new system is the capability of performing stationary counts to plot vertical profiles of iodine content through any portion of the gland. This new system should be very useful for studies designed to differentiate between malignant and benign thyroid nodules on the basis of iodine content.
Keywords
Collimators; Detectors; Fluorescence; Glands; High-resolution imaging; Image resolution; Optical imaging; Sensor arrays; Tomography; X-ray imaging;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1980.4330863
Filename
4330863