  • Title

    Corrections for Attenuation, Scattered Radiation, and Random Coincidences in a Ring Detector Positron Emission Transaxial Tomograph

  • Author

    Bergstrom, M ; Bohm, C. ; Ericson, K ; Eriksson, L. ; Litton, J.

  • Author_Institution
    Departments of Neuroradiology and Neurophysiology, Karolinska Hospital, Department of Radiation Physics, Karolinska Institute, and Department of Physics, University of Stockholm, Stockholm, Sweden
  • Volume
    27
  • Issue
    1
  • fYear
    1980
  • Firstpage
    549
  • Lastpage
    554
  • Abstract
    The effect of attenuation, scattered radiation and random coincidences were evaluated in a positron camera with 95 NaI-detectors. Experimental tests and computer simulations were performed to find the contribution of the various effects to the number of counts in each detector channel. Images were then reconstructed with and without corrections for the different factors and the errors in the images were quantified. It is shown that accurate quantitative measurements of regional concentration of radionuclides necessitate a proper correction for the different factors.