• DocumentCode
    849030
  • Title

    Data Acquisition and Processing Electronics for Positron Computed Tomography

  • Author

    Burgiss, S.G. ; Williams, C.W.

  • Volume
    27
  • Issue
    1
  • fYear
    1980
  • Firstpage
    649
  • Lastpage
    654
  • Abstract
    Positron emission computed axial tomography requires very accurate coincidence measurements, with large numbers of detector elements and large data arrays, for good image quality with high signal-to-noise (S/N) ratios. This paper presents the design concepts and system implementation approaches used to attain a system time resolution of 3.8 nsec and 7.6 nsec FWHM and FWTM, respectively, on a system using 66 NaI(Tl) detectors, 3.8cm diameter by 7.6cm long. Additionally, the system response provides a data processing burst rate of >5 MHz and dynamic random count subtraction without additional dead-time losses on a true line-of-response basis.
  • Keywords
    Bismuth; Circuits; Computed tomography; Data acquisition; Delay; Detectors; Event detection; Histograms; Positrons; Sensor arrays;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1980.4330907
  • Filename
    4330907