• DocumentCode
    851335
  • Title

    A Portable Device for the Measurement of Regional Cerebral Blood Flow in the Icu and or Using Cd Te Detectors and a Fourier Transform Based Data Analysis

  • Author

    Correia, J.A. ; Ackerman, R.H. ; Buonanno, F. ; Kaufman, D. ; Skiver, J. ; Alpert, N. ; Taveras, J. ; Entine, G.

  • Author_Institution
    Massachusetts General Hospital, Boston, MA 02114
  • Volume
    28
  • Issue
    1
  • fYear
    1981
  • Firstpage
    50
  • Lastpage
    54
  • Abstract
    A portable system for the measurement of cerebral blood flow in critical care environments such as the intensive care unit or operating room has been developed and tested in a limited series of patients. The flow measurements are based on the 133Xe inhalation or IV injection method. The system uses miniature solid state detectors, microprocessor technology and Fourier transform based data analysis to provide compactness and rapid on-line analysis of data. Preliminary patient studies comparing this system with a larger sodium iodide detector system show comparable results.
  • Keywords
    Blood flow; Data analysis; Fluid flow measurement; Fourier transforms; Hospitals; Microprocessors; Radiation detectors; Solid state circuits; System testing; Tellurium;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1981.4331138
  • Filename
    4331138