DocumentCode
851335
Title
A Portable Device for the Measurement of Regional Cerebral Blood Flow in the Icu and or Using Cd Te Detectors and a Fourier Transform Based Data Analysis
Author
Correia, J.A. ; Ackerman, R.H. ; Buonanno, F. ; Kaufman, D. ; Skiver, J. ; Alpert, N. ; Taveras, J. ; Entine, G.
Author_Institution
Massachusetts General Hospital, Boston, MA 02114
Volume
28
Issue
1
fYear
1981
Firstpage
50
Lastpage
54
Abstract
A portable system for the measurement of cerebral blood flow in critical care environments such as the intensive care unit or operating room has been developed and tested in a limited series of patients. The flow measurements are based on the 133Xe inhalation or IV injection method. The system uses miniature solid state detectors, microprocessor technology and Fourier transform based data analysis to provide compactness and rapid on-line analysis of data. Preliminary patient studies comparing this system with a larger sodium iodide detector system show comparable results.
Keywords
Blood flow; Data analysis; Fluid flow measurement; Fourier transforms; Hospitals; Microprocessors; Radiation detectors; Solid state circuits; System testing; Tellurium;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1981.4331138
Filename
4331138
Link To Document