Title :
Fast, High Flux, Photovoltaic Cdte Detector
Author :
Entine, G. ; Squillante, M.R. ; Serreze, H.B. ; Clarke, E.
Author_Institution :
Radiation Monitoring Devices, Inc. 44 Hunt Street Watertown, MA 02172
Abstract :
A recently invented CdTe radiation detector designed for high flux applications has been improved to give excellent response over a wide range of x-ray and gamma-ray fluxes. This device operates without an externally applied bias and has no dark current, which makes it very sensitive even at low flux. Measurements over a range of 10-1R/hr to 106 R/hr (Flux of ~ 104 to 1011 photons/cm2-sec) show both nearly linear response and high signal-to-noise. Afterglow has been reduced dramatically so that 1 ms time response can be achieved. This advance opens the possibility of using this semiconductor in the "current mode" of operation for both medical and industrial applications.
Keywords :
Dark current; Electrical equipment industry; Instruments; Photoconductivity; Photovoltaic cells; Photovoltaic systems; Radiation detectors; Semiconductor materials; Semiconductor radiation detectors; Solar power generation;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1981.4331239