    Performance Studies of High Gain Photomultiplier Having Z-Configuration of Microchannel Plates

  • Author

    Lo, C.C. ; Leskovar, B.

  • Author_Institution
    Lawrence Berkeley Laboratory University of California Berkeley, California 94720 U.S.A.
  • Volume
    28
  • Issue
    1
  • fYear
    1981
  • Firstpage
    698
  • Lastpage
    704
  • Abstract
    The characteristics of a high gain type ITT F4129 photomultiplier having three microchannel plates in cascade for electron multiplications have been investigated. These plates are in the Z-configuration. Measurements are given of the gain, dark current, cathode quantum efficiency, anode pulse linearity, electron transit time, single and multiphoton time spreads, fatigue, and pulse height resolution. The gain as a function of transverse magnetic field has been measured and is discussed. Photomultiplier characteristics as a function of the input pulse repetition frequency have also been investigated and discussed.
  • Keywords
    Current measurement; Dark current; Electrons; Gain measurement; Magnetic field measurement; Microchannel; Performance gain; Photomultipliers; Pulse measurements; Time measurement;