Comparison between a germanium orthogonal strip detector and an Anger camera through a simulation and modeling study
Author
Gombia, Mirko ; Brill, Aaron B. ; Bollini, Dante ; Del Guerra, Alberto
Author_Institution
Environ. Control Eng. Dept., Bologna Univ., Italy
Volume
49
Issue
5
fYear
2002
fDate
10/1/2002 12:00:00 AM
Firstpage
2196
Lastpage
2202
Abstract
Progress in detector technology has led to the development of a new generation of pixel-based imaging devices as potential competitors for Anger-type scintillation cameras. We modeled a 11-mm-thick germanium orthogonal strip detector (GOSD) with 2-mm pitch and compared its performance to an Anger camera with a 10-mm-thick NaI(Tl) crystal. The Anger camera simulation method was validated by experimental measurements made with point and volume sources. Resolution and sensitivity were determined for air and scatter measurements. The device is intended for use in breast tumor imaging, and its expected performance was simulated in response to 5-, 7.5-, and 10-mm spherical tumors embedded in a previously reported phantom geometry used for the simulation of a CsI pixellated detector. Thorax, breast, and heart background contributions are considered in this phantom. A comparison of the results obtained indicates that the GOSD provides superior contrast than the Anger camera for every tumor to collimator distance and for every tumor dimension. Also, signal-to-noise ratio (SNR), full-width at half-maximum (FWHM), and full-width at tenth maximum (FWTM) show a better response for the germanium pixellated detector with respect to Anger camera due to better energy and spatial resolution of the germanium-based device.
Keywords
Monte Carlo methods; germanium radiation detectors; radioisotope imaging; solid scintillation detectors; 10 mm; 11 mm; 5 mm; 7.5 mm; Anger camera; FWHM; FWTM; Ge; Ge orthogonal strip detector; Monte Carlo; NaI:Tl; breast; energy resolution; gamma camera; heart; resolution; sensitivity; signal-to-noise ratio; spatial resolution; thorax; tumor; Breast neoplasms; Cameras; Germanium; Imaging phantoms; Pixel; Scattering; Solid modeling; Solid scintillation detectors; Strips; Volume measurement;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2002.803821
Filename
1046810
Link To Document