An accurate electric field strength for the frequency range of 30 MHz to 1000 MHz has been developed in a full anechoic chamber with 1-Σ uncertainty less than ±1.9%. Reflection parameters of walls at frequencies below 200 MHz are evaluated experimentally, and the errors due to these waves are compensated in the standard and calibrations of the antennas
Keywords
anechoic chambers; electromagnetic wave reflection; field strength measurement; measurement standards; 30 to 1000 MHz; anechoic chamber; calibrations; electric field strength; walls reflection parameters; Anechoic chambers; Antenna measurements; Antenna theory; Frequency; Measurement standards; Receiving antennas; Reflection; Reflector antennas; Standards development; Voltage;