DocumentCode :
872949
Title :
Cross Talk in Multielectrode Particle Detectors
Author :
Richer, J.P. ; Chase, R.L.
Author_Institution :
Laboratoire de l´´Accélérateur Linéaire, Orsay
Volume :
31
Issue :
1
fYear :
1984
Firstpage :
258
Lastpage :
263
Abstract :
In highly segmented particle detectors, the crosstalk coupling capacitance between segments may vary from several per cent to as much as 30 % of the total electrode capacitance. Techniques for reducing associated errors by using flat-topped filter weighting functions and low input inpedance preamplifiers will be presented. The damping of resonances due to lead inductance and transformer leakage inductance, which strongly affect the crosstalk signals, will be discussed. Practical circuits using sample-and-hold techniques will be presented.
Keywords :
Capacitance; Circuits; Crosstalk; Damping; Electrodes; Filters; Inductance; Preamplifiers; Radiation detectors; Resonance;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1984.4333258
Filename :
4333258
Link To Document :
بازگشت