• DocumentCode
    873684
  • Title

    Dynamic Imaging with High Resolution Time-of-Flight PET Camera - TOFPET I

  • Author

    Mullani, N.A. ; Gaeta, J. ; Yerian, K. ; Wong, W.-H. ; Hartz, R.K. ; Philippe, E.A. ; Bristow, D. ; Gould, K.L.

  • Author_Institution
    University of Texas Health Science Center Division of Cardiology 6431 Fannin, Houston, Texas 77030
  • Volume
    31
  • Issue
    1
  • fYear
    1984
  • Firstpage
    609
  • Lastpage
    613
  • Abstract
    One of the major design goals of the TOFPET I positron camera was to produce a high resolution whole body positron camera capable of dynamically imaging an organ such as the heart. TOFPET I is now nearing completion and preliminary images have been obtained to assess its dynamic and three dimensional imaging capabilities. Multiple gated images of the uptake of Rubidium in the dog heart and three dirensional surface displays of the distribution of the Rubidium-82 in the myocardium have been generated to demnstrate the three dimensional imaging properties. Fast dynamic images of the first pass of a bolus of radio-tracer through the heart have been collected with 4 second integration time and 50% gating (2 second equivalent integration time) with 18 mCi of Rb-82.
  • Keywords
    Cameras; Detectors; Heart; High-resolution imaging; Image reconstruction; Image resolution; Image sampling; Microprocessors; Positron emission tomography; Signal resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1984.4333329
  • Filename
    4333329