Abstract
One of the major design goals of the TOFPET I positron camera was to produce a high resolution whole body positron camera capable of dynamically imaging an organ such as the heart. TOFPET I is now nearing completion and preliminary images have been obtained to assess its dynamic and three dimensional imaging capabilities. Multiple gated images of the uptake of Rubidium in the dog heart and three dirensional surface displays of the distribution of the Rubidium-82 in the myocardium have been generated to demnstrate the three dimensional imaging properties. Fast dynamic images of the first pass of a bolus of radio-tracer through the heart have been collected with 4 second integration time and 50% gating (2 second equivalent integration time) with 18 mCi of Rb-82.
Keywords