• DocumentCode
    873908
  • Title

    Development of a Portable Microdosimetric Radiation Protection Monitor Covering a Dynamic Range of 120 dB above Noise

  • Author

    Arbel, A. ; Booz, J. ; Mhuller, K.D. ; Neuhaub, H.

  • Author_Institution
    KFA Jiulich GmbH, on leave of absence from the Israel Institute of Technology, Haifa, Israel
  • Volume
    31
  • Issue
    1
  • fYear
    1984
  • Firstpage
    691
  • Lastpage
    696
  • Abstract
    A portable microdosimeter has been designed, covering a dynamic range of 120 dB in a single signal processing channel. A transistor is employed as non-linear element and deviation at low and high energies from a truly logarithmic characteristic is corrected by a microcomputer. A description of the basic design and experimental result are given.
  • Keywords
    Biomedical signal processing; Dynamic range; Employment; Instruments; Microcomputers; Preamplifiers; Protection; Radiation detectors; Radiation monitoring; Signal design;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1984.4333348
  • Filename
    4333348