• DocumentCode
    878042
  • Title

    New Beam Diagnostics at GANIL : Very Sensitive Current Transformers in Beam Lines and Counting System of Beam Turns in Cyclotrons

  • Author

    Loyer, F. ; Andre, T. ; Ducoudret, B. ; Rataud, J.P.

  • Author_Institution
    GANIL, BP 5027, CAEN Cedex - France
  • Volume
    32
  • Issue
    5
  • fYear
    1985
  • Firstpage
    1938
  • Lastpage
    1940
  • Abstract
    Two new beam diagnostics have been designed and are in operation at GANIL : - To measure very small beam currents without intercepting the beam, transformers have been developped whose resolution is as low as 1 enA. That is obtained by ussing synchronous detectors locked on the PIG source pulsation and by reducing the noise level. - As the accelerated orbits in the second main cyclotron are overlapping one another, we cannot count the number of turns by a classical moving probe. Therefore we compute the cross-correlation function of the beam signals detected on inductive probes at the input and the output of the cyclotron. Due to the natural random beam structure, this cross-correlation function has a sharp maximum for the value of the beam time of flight. Thence the right number of beam turns can be calculated.
  • Keywords
    Acceleration; Current measurement; Current transformers; Cyclotrons; Detectors; Extraterrestrial measurements; Noise level; Orbits; Probes; Signal detection;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1985.4333773
  • Filename
    4333773