DocumentCode
878112
Title
Frequency domain multiplexing for parallel acquisition of MR images
Author
Xu, Q. ; Wang, H. ; Xu, Z. ; Li, G.
Author_Institution
Dept. of Phys., East China Normal Univ., Shanghai, China
Volume
42
Issue
6
fYear
2006
fDate
3/16/2006 12:00:00 AM
Firstpage
326
Lastpage
327
Abstract
A frequency domain multiplexing (FDM) technique is described for parallel acquisition of magnetic resonance images. Two signals from different RF coils encoded with different frequencies were combined to feed into a high-speed ADC. The multiplexed signal is then decoded by a commercial multi-carrier digital receiver. Experimentation found no phase shift, no SNR loss and essentially no crosstalk in the proposed FDM method. It is expected that the FDM technique, with the development of commercially available digital receivers, may be valuable and cost-effective in the design of a parallel MRI receiver system.
Keywords
analogue-digital conversion; biomedical MRI; frequency division multiplexing; frequency-domain analysis; medical image processing; receivers; MR images; MRI receiver system; RF coils; frequency domain multiplexing; high-speed analog-to-digital converter; magnetic resonance images; multicarrier digital receiver; multiplexed signal; parallel acquisition;