• DocumentCode
    878112
  • Title

    Frequency domain multiplexing for parallel acquisition of MR images

  • Author

    Xu, Q. ; Wang, H. ; Xu, Z. ; Li, G.

  • Author_Institution
    Dept. of Phys., East China Normal Univ., Shanghai, China
  • Volume
    42
  • Issue
    6
  • fYear
    2006
  • fDate
    3/16/2006 12:00:00 AM
  • Firstpage
    326
  • Lastpage
    327
  • Abstract
    A frequency domain multiplexing (FDM) technique is described for parallel acquisition of magnetic resonance images. Two signals from different RF coils encoded with different frequencies were combined to feed into a high-speed ADC. The multiplexed signal is then decoded by a commercial multi-carrier digital receiver. Experimentation found no phase shift, no SNR loss and essentially no crosstalk in the proposed FDM method. It is expected that the FDM technique, with the development of commercially available digital receivers, may be valuable and cost-effective in the design of a parallel MRI receiver system.
  • Keywords
    analogue-digital conversion; biomedical MRI; frequency division multiplexing; frequency-domain analysis; medical image processing; receivers; MR images; MRI receiver system; RF coils; frequency domain multiplexing; high-speed analog-to-digital converter; magnetic resonance images; multicarrier digital receiver; multiplexed signal; parallel acquisition;
  • fLanguage
    English
  • Journal_Title
    Electronics Letters
  • Publisher
    iet
  • ISSN
    0013-5194
  • Type

    jour

  • DOI
    10.1049/el:20063890
  • Filename