Abstract
Transmit B 1 + field homogenization in high-field ( > 3.0 T) human magnetic resonance imaging (MRI) is challenging due to radio-frequency wavelength effects. An approach based on appropriately coupling surface coils to a volume coil was investigated. Electromagnetic simulation results demonstrated the feasibility and effectiveness of this method in proton MRI of the human head at 7.0 T.
Keywords
biomedical MRI; coils; B1 + field homogenization; electromagnetic simulation; high-field magnetic resonance imaging; human head; multilayer coupled coils; proton MRI; radio-frequency wavelength effects; Coils; Couplings; Humans; Magnetic heads; Magnetic multilayers; Magnetic resonance imaging; Molecular imaging; Protons; Radio control; Radio frequency; Magnetic resonance imaging (MRI); radio frequency coils; Algorithms; Electromagnetic Fields; Head; Humans; Magnetic Resonance Imaging; Models, Theoretical; Phantoms, Imaging;