• DocumentCode
    897807
  • Title

    A Hybrid Emission CT - HEADTOME II

  • Author

    Hirose, Y. ; Ikeda, Y. ; Higashi, Y. ; Koga, K. ; Hattori, H. ; Kanno, I. ; Miura, Y. ; Miura, S. ; Uemura, K.

  • Author_Institution
    R/D Engineering Department, Medical Systems Division, Shimadzu Corporation, Kyoto, Japan
  • Volume
    29
  • Issue
    1
  • fYear
    1982
  • Firstpage
    520
  • Lastpage
    523
  • Abstract
    HEADTOME II for a hybrid emission computed tomography for brain has been developed. The system includes multislice capability of both positron emission tomography (PET) and single photon emission tomography (SET) in one system. It consists of three detector rings of 42cm in diameter on which 64 NaI crystals are arranged per ring, 192 as a total. Four kinds of collimators are prepared: high resolution (HR) collimator and high sensitivity (HS) collimator for both SET and PET. The tangential spatial resolution is 6.5mm-11mm for SET with the HR collimator and 10.5mm-20mm with the HS collimator measured with 99mTc line source. System sensitivity with the single photon HS collimator is 58.5 kcps/¿Ci/ml for a 20cm diameter uniform solution of 99mTc. System sensitivity with the positron HS collimator is 25 kcps/¿Ci/ml for 68Ga solution.
  • Keywords
    Apertures; Computed tomography; Crystals; Detectors; Electrical capacitance tomography; Optical collimators; Positron emission tomography; Prototypes; Single photon emission computed tomography; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1982.4335899
  • Filename
    4335899