    A Multi-Element Silicon Detector for X-Ray Flux Measurements

  • Author

    Thompson, A.C. ; Goulding, F.S. ; Sommer, H.A. ; Walton, J.T. ; Hughes, E.B. ; Rolfe, J. ; Zeman, H.D.

  • Author_Institution
    Lawrence Berkeley Laboratory University of California Berkeley, California 94720 U.S.A.
  • Volume
    29
  • Issue
    1
  • fYear
    1982
  • Firstpage
    793
  • Lastpage
    797
  • Abstract
    A 30-element Si(Li) detector has been fabricated to measure the one-dimensional flux profile of 33 KeV x-rays from a synchrotron radiation beam. The device, which is fabricated from a single 39mm × 15mm silicon wafer, is a linear array of 0.9mm × 7mm elements with a 1mm center-to-center spacing. It is 5mm thick and when operated at room temperature has an average leakage current of 10 nA/element. The x-ray flux in each element is determined by measuring the current with a high quality operational amplifier followed by a current digitizer. This detector is being used to study the use of synchrotron radiation for non-invasive imaging of coronary arteries. The experiment uses the difference in the transmitted flux of a monochromatized x-ray beam above and below the iodine K-edge. Measurements have been made on plastic phantoms and on excised animal hearts with iodinated arteries. The images obtained indicate that a 256-element device with similar properties, but with 0.6mm element spacing, will make a very effective detector for high-speed medical imaging.
  • Keywords
    Arteries; Current measurement; Leakage current; Radiation detectors; Silicon; Synchrotron radiation; Temperature; X-ray detection; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1982.4335961
  • Filename
    4335961