• DocumentCode
    900445
  • Title

    Possible Bio-Medical Applications of Antiprotons I. In-Vivo Direct Density Measurements: Radiography

  • Author

    Gray, L. ; Kalogeropoulos, T.E.

  • Author_Institution
    Department of Physics, Syracuse University, Syracuse, New York 13210
  • Volume
    29
  • Issue
    2
  • fYear
    1982
  • fDate
    4/1/1982 12:00:00 AM
  • Firstpage
    1051
  • Lastpage
    1057
  • Abstract
    We show that antiprotons can be used to extract density at any point within a biological medium by measuring the stopping power. There are no theoretical limits to the accuracy of such a measurement, but practical ones come from data collection capabilities. In addition, radiation dosage may result in a limit of ¿¿/¿ = 10-2 when the measurements are taken within a volume of approximately 10-2 cm3. Measurements over larger volumes result in substantially smaller errors. The method is illustrated with computer simulated "experiments".
  • Keywords
    Attenuation measurement; Data mining; Density measurement; Mesons; Particle beams; Power measurement; Protons; Radiation dosage; Radiography; Tomography;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1982.4336313
  • Filename
    4336313