DocumentCode :
900700
Title :
Time Subtraction Digital Fluoroscopy with a Dual Memory System
Author :
Gould, Robert G.
Author_Institution :
Department of Radiology, University of California, San Francisco, California 94143
Volume :
29
Issue :
3
fYear :
1982
fDate :
6/1/1982 12:00:00 AM
Firstpage :
1186
Lastpage :
1190
Abstract :
This paper describes an x-ray fluoroscopic system for imaging arterial structures following intravenous injection of contrast medium. This system digitally processes video data from a TV camera that views the output of an x-ray image intensifier. The digital processing includes subtraction and contrast amplification. The unit (1) produces a real-time, continuous subtraction sequence, (2) improves the signal-to-noise ratio by averaging together up to 128 video frames generated at fluoroscopic levels, (3) does not require an interface to eh x-ray generator, and (4) can use either a moving or frozen mask image for subtraction.
Keywords :
Angiography; Arteries; Catheters; Digital cameras; Diseases; Image intensifiers; Optical imaging; Signal generators; TV; X-ray imaging;