DocumentCode
905374
Title
An Extended Range Well Counter
Author
Nickles, R.J. ; Votaw, J.R.
Author_Institution
Medical Physics Dept. 1530 MSC, 1300 University Ave, Madison, WI 53706
Volume
32
Issue
1
fYear
1985
Firstpage
60
Lastpage
62
Abstract
A well-type detector is described for the assay of gamma emitters over a Curie-to-nanoCurie activity range. A large plastic scintillator/photomultiplier is DC-coupled for current measurements at high activities, with pulse counting covering the low activities. The two techniques overlap for two decades in the uCi region, particularly troublesome for the alternative detectors: gas-filled ionization chambers or NaI(Tl) well counters. The plastic well is interfaced to a microcomputer for data acquisition and control of the phototube high voltage, the key to the nine-decade dynamic range.
Keywords
Counting circuits; Current measurement; Gamma ray detection; Gamma ray detectors; Ionization chambers; Microcomputers; Photomultipliers; Plastics; Pulse measurements; Radiation detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1985.4336789
Filename
4336789
Link To Document