DocumentCode
905421
Title
Sensitive Null Detector for Precise Measurements of Integral Linearity
Author
Takeuchi, S. ; Okumura, T.
Author_Institution
Department of Electronic Engineering, Saitama University 255 Shimo-Ohkubo, Urawa, Saitama, Japan
Volume
32
Issue
1
fYear
1985
Firstpage
74
Lastpage
77
Abstract
A method to increase the sensitivity of null detection in the linearity testing bridge is described. It is based on signal averaging to reduce the magnitude of noise in the error signal of the bridge output. Error signal of less than 0.1 mV(0.001% of 10 V) can easily be detected.
Keywords
Bridge circuits; Circuit noise; Circuit testing; Detectors; Linearity; Noise reduction; Pulse amplifiers; Sampling methods; Timing; Voltage;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1985.4336793
Filename
4336793