DocumentCode
907115
Title
Analysis of Positron-Emission Tomography Scintillation-Detectors with Wedge Faces and Inter-Crystal Septa
Author
Holmes, Timothy J. ; Ficke, David C.
Author_Institution
Mallinckrodt Institute of Radiology, and Institute for Biomedical Computing, Washington University School of Medicine, St. Louis, Missouri 63110
Volume
32
Issue
1
fYear
1985
Firstpage
826
Lastpage
830
Abstract
Various high-resolution scintillation detectors are analyzed for resolution and sensitivity. In particular, the effects of using a wedge face and inter-crystal tungsten-septa are studied for both BaF2 and BGO crystals. The results indicate that usage of septa alone causes an improvement in resolution in areas away from the center of the image, but also causes a degradation in sensitivity in those areas, and that usage of a wedge face alone causes no improvement in resolution, while usage of septa and a wedge face in combination causes a similar resolution improvement, but with a lesser degradation in sensitivity.
Keywords
Crystals; Degradation; Detectors; Face detection; Geometry; Image resolution; Scintillation counters; Spatial resolution; Tomography; Tungsten;