• DocumentCode
    907265
  • Title

    Quantitative Imaging with the MGH Analog Ring Positron Tomograph

  • Author

    Stearns, C.W. ; Chesler, D.A. ; Kirsch, J.E. ; Brownell, G.L.

  • Author_Institution
    Physics Research Laboratory Massachusetts General Hospital
  • Volume
    32
  • Issue
    1
  • fYear
    1985
  • Firstpage
    898
  • Lastpage
    901
  • Abstract
    Performance of the MGH positron camera, PCR-I, is evaluated at high count rates. System resolution is maintained throughout the operating range of the instrument. The analysis of dead-time losses and random coincidence events is complicated by the analog coding system implemented in PCR-I. However, the system non-idealities can be analyzed, and corrections applied to compensate for them. An experiment demonstrating the efficacy of the count rate corrections is presented.
  • Keywords
    Cameras; Error correction; Event detection; High-resolution imaging; Image reconstruction; Image resolution; Instruments; Positrons; Solid scintillation detectors; Wire;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1985.4336963
  • Filename
    4336963