Title :
A Curved Razor-Blade Proportional Counter for X-Ray Diffraction
Author :
Duijn, J.H. ; van Eijk, C.W.E. ; Hollander, R.W. ; Marx, R.
Author_Institution :
Physics Department, Delft University of Technology, Lorentzweg 1, 2600 GA DELFT, The Netherlands
Abstract :
A curved one-dimensional position-sensitive proportional detector (PSD) has been developed for powder diffraction studies. A curved "razor-blade" functions as anode. Cathode-strip readout gives position sensitivity. With individual strip readout, XeCO2 and Cu K¿ X-rays we obtained for the width of a diffraction line (FWHM) 200 ¿m (¿2¿ ¿ 0.1°).
Keywords :
Anodes; Blades; Cameras; Cathodes; Counting circuits; Detectors; Geometry; Powders; Strips; X-ray diffraction;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1986.4337126