  • Abstract
    Two possible detector designs for high resolution positron camera systems have been investigated. The goal is to achieve an instrument that can measure the whole brain with a spatial resolution of 5 mm FWHM in all directions. For both detectors BGO crystals are used, with the dimension 4.5*9.5*25 mm. One detector scheme utilizes the Anger principle for crystal identification with 16 crystals mounted on two dual PMT:s via a 3 mm light guide. The other detector scheme utilizes position sensitive PMT:s. The figures of merit for these two configurations are discussed in terms of high count rate capabilities and identification reliability.
  • Keywords
    Anodes; Automatic control; Cameras; Collimators; Crystals; Event detection; Logic; Position sensitive particle detectors; Positron emission tomography; Spatial resolution;