A characterization of the scatter point-spread-function in terms of air gaps
Author :
Wagner, Frederick C. ; Macovski, Albert ; Nishimura, Dwight G.
Author_Institution :
Inf. Syst. Lab., Stanford Univ., CA, USA
Volume :
7
Issue :
4
fYear :
1988
Firstpage :
337
Lastpage :
344
Abstract :
Characterizing the scatter point-spread-function (PSF) yields a model for predicting the behavior of the scatter and a PSF for deconvolution techniques that correct for scatter. Assuming the X-ray scatter is isotropic, the authors present a model for the scatter point-spread-function parameterized only by air gap. This model suggests that small increases in air gap significantly attenuate the higher-frequency structure of the scatter distribution. To evaluate this model, the authors examined the behavior of the spatial frequency distributions of experimental scatter images as a function of air gap. Using film as the detector, they imaged a 20-cm uniformly thick water phantom with aperture diameters of 8, 12, 16, 20, and 24 mm at air gaps of 0-24 cm.<>
Keywords :
X-ray scattering; diagnostic radiography; X-ray scatter; air gaps; aperture diameter; film; isotropic scattering; scatter point-spread-function; water phantom; Acoustic scattering; Air gaps; Apertures; Deconvolution; Frequency; Optical imaging; Predictive models; Scattering parameters; X-ray imaging; X-ray scattering;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.14517