DocumentCode :
934087
Title :
Performance Study of the New Hamamatsu R9779 and Photonis XP20D0 Fast 2” Photomultipliers
Author :
Bauer, Florian ; Aykac, Mehmet ; Loope, Michael ; Williams, Charles W. ; Eriksson, Lars ; Schmand, Matthias
Volume :
54
Issue :
3
fYear :
2007
fDate :
6/1/2007 12:00:00 AM
Firstpage :
422
Lastpage :
426
Abstract :
The focus of this paper is the evaluation of the new fast 51 mm-diameter, 8-stage Hamamatsu R9779 photomultipliers (PMTs) with an acceleration-ring at the front-end and the Photonis XP20D0 PMTs with a screening grid in front of the anode. The following performance characteristics are presented: Timing resolution, anode-scan-uniformity and transit-time spread. The unfolded timing resolution for two R9779 was 192 ps and 210 ps using plastic scintillators. The individual timing resolutions for two XP20D0 using plastic scintillators are 181 ps and 154 ps, respectively. The variation in time resolution across the windows of the two R9779 ranged between 117 ps and 171 ps, and 79 ps and 73 ps for the two XP20D0 PMTs.
Keywords :
anodes; photocathodes; photomultipliers; scintillation; timing; Hamamatsu R9779 photomultipliers; Photonis XP20D0 photomultipliers; acceleration-ring; anode-scan-uniformity; plastic scintillators; screening grid; size 51 mm; time 117 ps to 171 ps; time 181 ps; time 192 ps; time 210 ps; time 73 ps to 79 ps; timing resolution; transit-time spread; Anodes; Biomedical imaging; Cathodes; Detectors; Focusing; Photomultipliers; Plastics; Positron emission tomography; Speech synthesis; Timing; PET; photomultiplier; time-of-flight;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2007.895507
Filename :
4237388
Link To Document :
بازگشت