    Automated blood sampling systems for positron emission tomography

  • Author

    Eriksson, L. ; Holte, S. ; Bohm, Chr ; Kesselberg, M. ; Hovander, B.

  • Author_Institution
    Dept. of Neuroradiol. & Neurophysiol., Karolinska Hospital/Inst., Stockholm, Sweden
  • Volume
    35
  • Issue
    1
  • fYear
    1988
  • Firstpage
    703
  • Lastpage
    707
  • Abstract
    An automated blood sampling system has been constructed and evaluated for use with positron-emission tomography (PET). The dispersion functions of two different detector units in the blood sampling system are compared. One is a plastic scintillator, and the other is a coincidence detector. No significant differences were found. Results from studies of blood-brain barrier transfer of a C-11 labelled receptor antagonist are discussed.<>
  • Keywords
    biomedical measurement; blood; computerised tomography; radioisotope scanning and imaging; PET; automated blood sampling system; blood-brain barrier transfer; coincidence detector; dispersion functions; plastic scintillator; positron emission tomography; Automatic control; Blood; Cameras; Catheters; Control systems; Plastics; Positron emission tomography; Sampling methods; Scintillation counters; Solid scintillation detectors;