DocumentCode
985386
Title
New in-situ calibration of diode detectors used in six-port network analyzers
Author
Colef, Gabriel ; Karmel, Paul R. ; Ettenberg, Morris
Author_Institution
Dept. of Electr. Eng., City Coll., City Univ. of New York, NY, USA
Volume
39
Issue
1
fYear
1990
fDate
2/1/1990 12:00:00 AM
Firstpage
201
Lastpage
204
Abstract
The authors present a novel method for calibrating the diode detectors used in six-port network analyzers without disconnecting the diodes. The method uses either a precision variable attenuator at the six-port input or a power meter connected to the six-port test port. No actual measurement of power is required to calibrate the diode detectors when an attenuator is used. The method is very well suited to in situ calibration of the diode detection circuits for the operation of the six-port with pulsed signals when the average power detected by the power meter drops to very low levels for low duty cycles. In these cases. The precision variable attenuator method is required
Keywords
attenuators; calibration; microwave reflectometry; network analysers; average power; diode detectors; in-situ calibration; power meter; precision variable attenuator; pulsed signals; six-port network analyzers; Attenuation measurement; Attenuators; Calibration; Circuits; Diodes; Envelope detectors; Intelligent networks; Power measurement; Testing; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.50444
Filename
50444
Link To Document